|
HOW TO PERFORM A
2-POINT CALIBRATION VERIFICATION ON YOUR VISCOMETER/RHEOMETER
Procedure for using regular spindles LV1-LV4, RV1-RV7, HA1-HA7 and
HB1-HB7; if you use any other spindles and/or accessories, or if you
have a special viscometer/rheometer, please
contact us.
Note: You cannot calibrate your viscometer as there is no
external provision for this, e.g. you cannot adjust the span or
zero; you can only verify the calibration.
What you
need:
- the to-be-checked viscometer
- the viscometer's guardleg (see NOTE below)
- a constant temperature bath
- one suitable Brookfield Standard Fluid
- a 600 ml Low Form Griffin beaker (see NOTE below)
NOTE: if the Brookfield Standard Fluid value is greater than
20,000 cps, the guardleg is not necessary for models LV and RV,
and the original viscosity standard fluid container can be used.
What to
do:
- pour the Brookfield Standard Fluid into the Griffin beaker
(see the above NOTE) and put the beaker in the bath;
- mount the viscometer on its stand beside the bath and level
it; if it is a digital, turn it on, autozero it and input the
spindle code;
- attach the spindle you intend to use and, if it's the case,
the guardleg;
- immerse to the appropriate depth the spindle + guardleg into
the Brookfield Standard Fluid;
- re-level the viscometer;
- turn the bath on and set it up to 25 degrees Celsius then
wait for at least 30 min (1 hour recommended) for the Brookfield
Standard Fluid temperature to stabilize;
- before taking a reading, recheck the viscometer leveling and
adjust if necessary;
- turn the viscometer on (dial) or the motor on (digital) and
take a high-torque % reading after at least 4 spindle rotations;
log reading details: speed, spindle, torque and viscosity;
- shift the speed down (typically half of the initial speed)
and allow the spindle to rotate for a few minutes; take a
lower-torque % reading and log it;
- calculate the accuracy ranges for both speeds - see below –
and determine if the readings are within calculated ranges; yes
means pass, no means fail.
Accuracy
range calculation:
The total error is a combined error Standard Fluid - Viscometer.
The accuracy for the Brookfield Standard Fluid is 1% of the actual
viscosity value written on the jar . The accuracy for the Viscometer
is 1% of the maximum range in use (note that is not 1% of the
reading!).
| EXAMPLE: Suppose we have a dial viscometer
RVT and a Brookfield Standard Fluid of 5000 cps with the
actual viscosity of 4,990 cps at 25 degrees Celsius. To
measure this we use an RV spindle #3 at the speed of 20 rpm.
The reading obtained is 4,970 cps. The error for the fluid is
1% of 4,990, that is 49.9 cps, and the error for the
viscometer is 1% of 5000 (maximum range in use given by the
spindle #3 at 20 rpm) , that is 50 cps. The total error is
49.9+50=99.9 cps. To calculate the range, substract and add
99.9 from 4990. The range is (4890.1 to 5089.9) cps. Any
reading within this range means the viscometer is
calibrated. In our example, 4,970 cps is within the above
range so the viscometer is OK. |
|
|

Mineral and Silicone Calibration Fluids
Click to enlarge

Silicone Calibration Fluids
Click to enlarge

Silicone Calibration Fluid with 600 mL Griffin Beaker
Click to enlarge
|