Can-Am Instruments Ltd. : Sales and service for laboratory and environmental equipment
Home > Support > Service/Repair and Calibration

Instrument Service/Repair and Calibration and Certification

  Instrument Calibration and Certification


Can-Am Instruments Ltd., an ISO 9001:2000 Certified company, is the exclusive dealer for sales and service in the territory of Canada for laboratory equipment sold by Brookfield Engineering Laboratories, Inc. We have been in business for over 45 years!

We also perform field service calls at customer's site for added convenience. Contact us for a quote to service your instrument(s) on site.

STANDARD CALIBRATION FOR A BROOKFIELD LABORATORY VISCOMETER

Note: This operation can only be performed by an authorized Brookfield representative

Whenever you send us your viscometer for re-calibration, or during a field service call, we will perform the following operations:

  • A general inspection: shaft straightness, motor and gear box (where applicable), pivot point and jewel bearing status;
  • A reading is taken and compared against our Master Viscometer; Before Calibration Data is collected;
  • The pivot point and pivot point support (known as jewel bearing) are replaced;
  • The unit is cleaned (including all bearings) and oiled/greased;
  • The unit is calibrated and tested;
  • A Certificate of Calibration is issued; the Certificate includes the Before and After Calibration Data;
  • Any accessories shipped with the unit are inspected, e.g. spindles for straightness.

We recommend re-calibration of the viscometer if one of the following is true:
- one year has elapsed since the last re-calibration;
- it fails the periodical calibration check using the Brookfield Standard fluid.

For more information or to arrange for your unit to be sent in, please call our Service Department at 905-829-0030 (Toll free 1-800-215-4469).

 



Instrument Calibration
Click to enlarge






 
  Instructions when sending Equipment to Us


When sending your equipment to us for service, please make sure that you read the instructions below and complete the attached Service/Calibration Request Form:

Service Instructions and Calibration Request Form (PDF)

 

 

 